HELP : DICOM initialization failure (version 2) - DICOM

This is a discussion on HELP : DICOM initialization failure (version 2) - DICOM ; Hi, I already revised my code based on comments and suggestions that I got in this forum. Here is the result dcm file. You can download it at www.materikuliah.com/DICOMTEST2.zip (it was zipped). Yes, this still can't be initialized by any ...

+ Reply to Thread
Page 1 of 2 1 2 LastLast
Results 1 to 20 of 22

Thread: HELP : DICOM initialization failure (version 2)

  1. HELP : DICOM initialization failure (version 2)

    Hi, I already revised my code based on comments and suggestions that I
    got in this forum. Here is the result dcm file. You can download it at
    www.materikuliah.com/DICOMTEST2.zip (it was zipped).

    Yes, this still can't be initialized by any DICOM viewers and tools.
    Please review it and tell me what's going wrong.

    Thank you for your comments, suggestions, and helps.

    Best regards,
    Budhi

  2. Re: HELP : DICOM initialization failure (version 2)

    The first data element after the preamble, the meta data group length, has
    a VR of UL (correct) and a VL of 4 (correct) but you have only encoded two
    bytes for the value itself, not 4.

    David

    budhik wrote:
    > Hi, I already revised my code based on comments and suggestions that I
    > got in this forum. Here is the result dcm file. You can download it at
    > www.materikuliah.com/DICOMTEST2.zip (it was zipped).
    >
    > Yes, this still can't be initialized by any DICOM viewers and tools.
    > Please review it and tell me what's going wrong.
    >
    > Thank you for your comments, suggestions, and helps.
    >
    > Best regards,
    > Budhi


  3. Re: HELP : DICOM initialization failure (version 2)

    By the way, Budhi, did you get my complete example the other day ?

    It posted successfully, but it may be too long for some news hosts,
    like Google.

    I put the text of the message, as well as the original DICOM file
    sample at:

    http://idisk.mac.com/dclunie-Public/ECG/

    David

    budhik wrote:
    > Hi, I already revised my code based on comments and suggestions that I
    > got in this forum. Here is the result dcm file. You can download it at
    > www.materikuliah.com/DICOMTEST2.zip (it was zipped).
    >
    > Yes, this still can't be initialized by any DICOM viewers and tools.
    > Please review it and tell me what's going wrong.
    >
    > Thank you for your comments, suggestions, and helps.
    >
    > Best regards,
    > Budhi


  4. Re: HELP : DICOM initialization failure (version 2)

    On Jul 18, 9:04 am, David Clunie wrote:
    > By the way, Budhi, did you get my complete example the other day ?
    >
    > It posted successfully, but it may be too long for some news hosts,
    > like Google.
    >
    > I put the text of the message, as well as the original DICOM file
    > sample at:
    >
    > http://idisk.mac.com/dclunie-Public/ECG/
    >
    > David
    >
    > budhik wrote:
    > > Hi, I already revised my code based on comments and suggestions that I
    > > got in this forum. Here is the result dcm file. You can download it at
    > >www.materikuliah.com/DICOMTEST2.zip(it was zipped).

    >
    > > Yes, this still can't be initialized by any DICOM viewers and tools.
    > > Please review it and tell me what's going wrong.

    >
    > > Thank you for your comments, suggestions, and helps.

    >
    > > Best regards,
    > > Budhi


    Hi David,

    I have downloaded the ecg.dcm and ecg.txt from your idisk folder. It
    can be read by using DVTK dicom editor, but when I tried to view it by
    using dicom ecg viewer (http://www.excel-medical.com/waveforms/Viewer/
    DicomWaveformViewer.htm), it can not be read. I don't know why.

    Btw, as I wrote my code based on your and other replier of comments in
    this group, I suspect that the problem is on the translation of VR's
    attributes into C's data type and the value length mis-understanding.
    For example, in VR of "US", the max length is 2 byte. But in C code, I
    should declare an unsigned short (since unsigned int can be 2 or 4
    bytes depend on our machine architecture; in my PC unsigned int will
    be represent as 4 bytes / 32 bit). Also for example, VR of "DS" with
    max length 16 bytes. I'm not sure whether it should be declare as
    char, unsigned char, or unsigned int, because the value can be
    numeric.

    So please help me to mapping these VRs and C data types :

    VR | Max Length | C Data type
    AE | 16 | .......
    AS | 4 | .......
    CS | 16 | .......
    DA | 8 | .......
    DS | 16 | .......
    IS | 12 | .......
    LO | 64 | .......
    OB | - | .......
    OW | - | .......
    PN | - | .......
    SH | 16 | .......
    SQ | - | .......
    ST | 1024 | .......
    TM | 16 | .......
    UI | 64 | .......
    UL | 4 | .......
    US | 2 | .......

    Another question :

    If I have (0008,0013) with VR of "TM" and value "165520", since "TM"
    max length is 16 bytes, should I declare in my C code unsigned short
    (numeric 16 bit / 2 byte) or unsigned char[16] or unsigned char[6] or
    what ? I'm so confuse about this mappings.

    Please advise,
    Budhi

  5. Re: HELP : DICOM initialization failure (version 2)

    On Jul 18, 3:51 pm, budhik wrote:
    > On Jul 18, 9:04 am, David Clunie wrote:
    >
    >
    >
    > > By the way, Budhi, did you get my complete example the other day ?

    >
    > > It posted successfully, but it may be too long for some news hosts,
    > > like Google.

    >
    > > I put the text of the message, as well as the original DICOM file
    > > sample at:

    >
    > >http://idisk.mac.com/dclunie-Public/ECG/

    >
    > > David

    >
    > > budhik wrote:
    > > > Hi, I already revised my code based on comments and suggestions that I
    > > > got in this forum. Here is the result dcm file. You can download it at
    > > >www.materikuliah.com/DICOMTEST2.zip(itwas zipped).

    >
    > > > Yes, this still can't be initialized by any DICOM viewers and tools.
    > > > Please review it and tell me what's going wrong.

    >
    > > > Thank you for your comments, suggestions, and helps.

    >
    > > > Best regards,
    > > > Budhi

    >
    > Hi David,
    >
    > I have downloaded the ecg.dcm and ecg.txt from your idisk folder. It
    > can be read by using DVTK dicom editor, but when I tried to view it by
    > using dicom ecg viewer (http://www.excel-medical.com/waveforms/Viewer/
    > DicomWaveformViewer.htm), it can not be read. I don't know why.
    >
    > Btw, as I wrote my code based on your and other replier of comments in
    > this group, I suspect that the problem is on the translation of VR's
    > attributes into C's data type and the value length mis-understanding.
    > For example, in VR of "US", the max length is 2 byte. But in C code, I
    > should declare an unsigned short (since unsigned int can be 2 or 4
    > bytes depend on our machine architecture; in my PC unsigned int will
    > be represent as 4 bytes / 32 bit). Also for example, VR of "DS" with
    > max length 16 bytes. I'm not sure whether it should be declare as
    > char, unsigned char, or unsigned int, because the value can be
    > numeric.
    >
    > So please help me to mapping these VRs and C data types :
    >
    > VR | Max Length | C Data type
    > AE | 16 | .......
    > AS | 4 | .......
    > CS | 16 | .......
    > DA | 8 | .......
    > DS | 16 | .......
    > IS | 12 | .......
    > LO | 64 | .......
    > OB | - | .......
    > OW | - | .......
    > PN | - | .......
    > SH | 16 | .......
    > SQ | - | .......
    > ST | 1024 | .......
    > TM | 16 | .......
    > UI | 64 | .......
    > UL | 4 | .......
    > US | 2 | .......
    >
    > Another question :
    >
    > If I have (0008,0013) with VR of "TM" and value "165520", since "TM"
    > max length is 16 bytes, should I declare in my C code unsigned short
    > (numeric 16 bit / 2 byte) or unsigned char[16] or unsigned char[6] or
    > what ? I'm so confuse about this mappings.
    >
    > Please advise,
    > Budhi


    Hi,

    I already mapped the VRs into C data type based on section 6.2 dicom
    spec file number 5.
    Based on that, I revised the code and generate new dcm file. This file
    shall contains ecg signals from the text file.
    You can see the dcm and txt file here (zipped) : www.materikuliah.com/DICOMTEST3.zip
    and www.materikuliah.com/12ecgsample2.zip

    The text file contains stream of 12 lead ecg signal which is separated
    into 14 column : time, ID, lead I, lead II, lead III, lead aVR, lead
    aVL, lead aVF, lead V1, lead V2, lead V3, lead V4, lead V5, lead V6.

    But still, the dcm file can't be initialized. I don't know why. I
    already compared my dcm file with 39da47b7.dcm (downloaded from
    http://www.excel-medical.com/waveforms/Waveform.htm), and I found the
    meta section seems the same. The only differences are waveform data
    and several info like patient name etc.

    Please give me further advises.

    Best regards,
    Budhi

  6. Sample ECG problems, was Re: HELP : DICOM initialization failure(version 2)

    Hi Budhi

    budhik wrote:

    >> I have downloaded the ecg.dcm and ecg.txt from your idisk folder. It
    >> can be read by using DVTK dicom editor, but when I tried to view it by
    >> using dicom ecg viewer (http://www.excel-medical.com/waveforms/Viewer/
    >> DicomWaveformViewer.htm), it can not be read. I don't know why.


    Actually there are two issues with this file that may have caused problems
    with that viewer:

    - the waveform bits allocated is > 8 but the explicit VR for the waveform
    data attribute is OB, which is illegal - my validator was not checking
    for that so I hadn't noticed

    - there are two multiplex groups, the first containing the 12 lead ECG and
    the second contained some derived waveforms

    I have corrected the file to use the OW VR, and the result is in this
    folder:

    http://idisk.mac.com/dclunie-Public/ECG/

    It displays fine with my own ECG viewer, which always uses only the first
    multiplex group and ignores the rest. See the screenshot at:

    http://idisk.mac.com/dclunie-Public/ECG/

    You can try the viewer at:

    http://www.dclunie.com/pixelmed/soft...ECGViewer.html

    David

  7. Re: HELP : DICOM initialization failure (version 2)

    Hi Budhi

    You cannot depend on the internal C or C++ data type lengths.

    You need to explicitly handle the number of bytes to be written
    and the order in which those bytes are written for all the binary
    DICOM VRs. The internal representation only has to be a) long
    enough, and b) handle sign extension properly if you are going
    to perform arithmetic or comparison on those values.

    All the string and text VRs can be handled as arrays of char, and
    padded to even length and the correct length written.

    That includes DA, TM, DT and DS ... those are all strings when
    encoded (regardless of whether they are stored "internally" as
    long or double or whatever).

    Why don't you just look at the source of one of the many C++ or
    Java toolkits ?

    David

    budhik wrote:

    > Btw, as I wrote my code based on your and other replier of comments in
    > this group, I suspect that the problem is on the translation of VR's
    > attributes into C's data type and the value length mis-understanding.
    > For example, in VR of "US", the max length is 2 byte. But in C code, I
    > should declare an unsigned short (since unsigned int can be 2 or 4
    > bytes depend on our machine architecture; in my PC unsigned int will
    > be represent as 4 bytes / 32 bit). Also for example, VR of "DS" with
    > max length 16 bytes. I'm not sure whether it should be declare as
    > char, unsigned char, or unsigned int, because the value can be
    > numeric.
    >
    > So please help me to mapping these VRs and C data types :
    >
    > VR | Max Length | C Data type
    > AE | 16 | .......
    > AS | 4 | .......
    > CS | 16 | .......
    > DA | 8 | .......
    > DS | 16 | .......
    > IS | 12 | .......
    > LO | 64 | .......
    > OB | - | .......
    > OW | - | .......
    > PN | - | .......
    > SH | 16 | .......
    > SQ | - | .......
    > ST | 1024 | .......
    > TM | 16 | .......
    > UI | 64 | .......
    > UL | 4 | .......
    > US | 2 | .......
    >
    > Another question :
    >
    > If I have (0008,0013) with VR of "TM" and value "165520", since "TM"
    > max length is 16 bytes, should I declare in my C code unsigned short
    > (numeric 16 bit / 2 byte) or unsigned char[16] or unsigned char[6] or
    > what ? I'm so confuse about this mappings.
    >
    > Please advise,
    > Budhi


  8. Re: HELP : DICOM initialization failure (version 2)

    budhik wrote:

    > Based on that, I revised the code and generate new dcm file. This file
    > shall contains ecg signals from the text file.
    > You can see the dcm and txt file here (zipped) : www.materikuliah.com/DICOMTEST3.zip
    > and www.materikuliah.com/12ecgsample2.zip

    ....
    > But still, the dcm file can't be initialized. I don't know why. I
    > already compared my dcm file with 39da47b7.dcm (downloaded from
    > http://www.excel-medical.com/waveforms/Waveform.htm), and I found the
    > meta section seems the same. The only differences are waveform data
    > and several info like patient name etc.


    % andump /Users/dclunie/Downloads/DICOMTEST3.dcm
    (0x0002,0x0000) UL File Meta Information Group Length VR=
      VL=<0x0004> [0x0000008c]
      (0x0002,0x0001) OB File Meta Information Version VR= VL=<0x0004> [0x01,0x00,0x02,0x00]
      (0x0002,0x4955) ? VR= VL=<0x2e31001e> [] # skipping ...

      % dciodvfy /Users/dclunie/Downloads/DICOMTEST3.dcm
      Warning - Bad File Meta Information Version Attribute - wrong VR, length or not 0x00,0x01
      (0x0002,0x4955) ? Error - Tag read failed - Implicit VR encoding even though supposed to be explicit
      Error - Dicom dataset read failed

  9. Re: HELP : DICOM initialization failure (version 2)

    On Jul 19, 4:31 am, David Clunie wrote:
    > budhik wrote:
    > > Based on that, I revised the code and generate new dcm file. This file
    > > shall contains ecg signals from the text file.
    > > You can see the dcm and txt file here (zipped) :www.materikuliah.com/DICOMTEST3.zip
    > > andwww.materikuliah.com/12ecgsample2.zip

    >
    > ...
    > > But still, the dcm file can't be initialized. I don't know why. I
    > > already compared my dcm file with 39da47b7.dcm (downloaded from
    > >http://www.excel-medical.com/waveforms/Waveform.htm), and I found the
    > > meta section seems the same. The only differences are waveform data
    > > and several info like patient name etc.

    >
    > % andump /Users/dclunie/Downloads/DICOMTEST3.dcm
    > (0x0002,0x0000) UL File Meta Information Group Length VR=
      VL=<0x0004> [0x0000008c]
      > (0x0002,0x0001) OB File Meta Information Version VR= VL=<0x0004> [0x01,0x00,0x02,0x00]
      > (0x0002,0x4955) ? VR= VL=<0x2e31001e> [] # skipping ...
      >
      > % dciodvfy /Users/dclunie/Downloads/DICOMTEST3.dcm
      > Warning - Bad File Meta Information Version Attribute - wrong VR, length or not 0x00,0x01
      > (0x0002,0x4955) ? Error - Tag read failed - Implicit VR encoding even though supposed to be explicit
      > Error - Dicom dataset read failed


      Hi,

      I don't understand how the mis-interpretation happened. As we can see,
      the bit values are the same.

      The 39da47b7.dcm :
      02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      and my DICOMTEST3.dcm :
      02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      How can my file was interpreted as (0002,4955) when the 39da47b7.dcm
      was interpreted as (0002,0002) and "UI" ???
      All the meta values are the same. The different things are : the value
      of date, patient name, and waveform signal.

      Is the DICOM waveform file should has the same end of bits of the
      file ?

      Please advise,
      Budhi

  10. Re: HELP : DICOM initialization failure (version 2)

    On Jul 19, 8:08 pm, budhik wrote:
    > On Jul 19, 4:31 am, David Clunie wrote:
    >
    >
    >
    > > budhik wrote:
    > > > Based on that, I revised the code and generate new dcm file. This file
    > > > shall contains ecg signals from the text file.
    > > > You can see the dcm and txt file here (zipped) :www.materikuliah.com/DICOMTEST3.zip
    > > > andwww.materikuliah.com/12ecgsample2.zip

    >
    > > ...
    > > > But still, the dcm file can't be initialized. I don't know why. I
    > > > already compared my dcm file with 39da47b7.dcm (downloaded from
    > > >http://www.excel-medical.com/waveforms/Waveform.htm), and I found the
    > > > meta section seems the same. The only differences are waveform data
    > > > and several info like patient name etc.

    >
    > > % andump /Users/dclunie/Downloads/DICOMTEST3.dcm
    > > (0x0002,0x0000) UL File Meta Information Group Length VR=
      VL=<0x0004> [0x0000008c]
      > > (0x0002,0x0001) OB File Meta Information Version VR= VL=<0x0004> [0x01,0x00,0x02,0x00]
      > > (0x0002,0x4955) ? VR= VL=<0x2e31001e> [] # skipping ...

      >
      > > % dciodvfy /Users/dclunie/Downloads/DICOMTEST3.dcm
      > > Warning - Bad File Meta Information Version Attribute - wrong VR, length or not 0x00,0x01
      > > (0x0002,0x4955) ? Error - Tag read failed - Implicit VR encoding even though supposed to be explicit
      > > Error - Dicom dataset read failed

      >
      > Hi,
      >
      > I don't understand how the mis-interpretation happened. As we can see,
      > the bit values are the same.
      >
      > The 39da47b7.dcm :
      > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49
      >
      > and my DICOMTEST3.dcm :
      > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49
      >
      > How can my file was interpreted as (0002,4955) when the 39da47b7.dcm
      > was interpreted as (0002,0002) and "UI" ???
      > All the meta values are the same. The different things are : the value
      > of date, patient name, and waveform signal.
      >
      > Is the DICOM waveform file should has the same end of bits of the
      > file ?
      >
      > Please advise,
      > Budhi


      Hi, I already upload the 4th result file in www.materikuliah.com/DICOMTEST4.zip
      (it was zipped).
      The good thing is the dcm file can be opened by using Dicom
      Inspector .Net , and I can see the wrong length of VR and the wrong
      value also. I already fix them based on David's ecg_ow.dcm file and
      39da47b7.dcm file.

      But, again, it can be initialized. I suspect the next wrong thing is
      in the waveform data syntax.

      Here is my (5400,1010) "OW" with 0x0000 of reserved and 0xffffffff of
      length (in C code) :

      ------- headers -----

      typedef struct specstruct {
      unsigned int ge;
      unsigned char vr[2];
      unsigned short reserved;
      unsigned long vl;
      };
      struct specstruct metaspec;

      float date,id,I,II,III,aVR,aVL,aVF,V1,V2,V3,V4,V5,V6;

      ------- main (bla... bla...) ----------

      fpecg = fopen("12ecgsample2.txt", "r");
      fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
      &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);

      metaspec.ge = 0x10105400;
      strcpy(metaspec.vr,"OW");
      metaspec.reserved = 0x0000;
      metaspec.vl = 0xffffffff;
      fwrite(&metaspec,sizeof(metaspec),1,fp);

      while (!feof(fpecg))
      {
      fwrite(&I,sizeof(I),1,fp);
      fwrite(&II,sizeof(II),1,fp);
      fwrite(&III,sizeof(III),1,fp);
      fwrite(&aVR,sizeof(aVR),1,fp);
      fwrite(&aVL,sizeof(aVL),1,fp);
      fwrite(&aVF,sizeof(aVF),1,fp);
      fwrite(&V1,sizeof(V1),1,fp);
      fwrite(&V2,sizeof(V2),1,fp);
      fwrite(&V3,sizeof(V3),1,fp);
      fwrite(&V4,sizeof(V4),1,fp);
      fwrite(&V5,sizeof(V5),1,fp);
      fwrite(&V6,sizeof(V6),1,fp);

      fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
      &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);
      }

      ---------- continue to the end --------------

      The signal value of each ecg channels are in float numbers %2.6f.

      Question :

      1. Is it correct to write the stream of signals without delimiter
      between signals ?
      2. Should it (the signals) converted first into hex before written
      into the file ?
      3. How to count the group length (0002,0000) value ?
      4. Is the (003a,0010) number of waveform samples should be exactly the
      same with the number of lines of signals in the text file ?

      Please advise,
      Budhi

  11. Re: HELP : DICOM initialization failure (version 2)

    On Jul 20, 1:14 am, budhik wrote:
    > On Jul 19, 8:08 pm, budhik wrote:
    >
    >
    >
    > > On Jul 19, 4:31 am, David Clunie wrote:

    >
    > > > budhik wrote:
    > > > > Based on that, I revised the code and generate new dcm file. This file
    > > > > shall contains ecg signals from the text file.
    > > > > You can see the dcm and txt file here (zipped) :www.materikuliah.com/DICOMTEST3.zip
    > > > > andwww.materikuliah.com/12ecgsample2.zip

    >
    > > > ...
    > > > > But still, the dcm file can't be initialized. I don't know why. I
    > > > > already compared my dcm file with 39da47b7.dcm (downloaded from
    > > > >http://www.excel-medical.com/waveforms/Waveform.htm), and I found the
    > > > > meta section seems the same. The only differences are waveform data
    > > > > and several info like patient name etc.

    >
    > > > % andump /Users/dclunie/Downloads/DICOMTEST3.dcm
    > > > (0x0002,0x0000) UL File Meta Information Group Length VR=
      VL=<0x0004> [0x0000008c]
      > > > (0x0002,0x0001) OB File Meta Information Version VR= VL=<0x0004> [0x01,0x00,0x02,0x00]
      > > > (0x0002,0x4955) ? VR= VL=<0x2e31001e> [] # skipping ...

      >
      > > > % dciodvfy /Users/dclunie/Downloads/DICOMTEST3.dcm
      > > > Warning - Bad File Meta Information Version Attribute - wrong VR, length or not 0x00,0x01
      > > > (0x0002,0x4955) ? Error - Tag read failed - Implicit VR encoding even though supposed to be explicit
      > > > Error - Dicom dataset read failed

      >
      > > Hi,

      >
      > > I don't understand how the mis-interpretation happened. As we can see,
      > > the bit values are the same.

      >
      > > The 39da47b7.dcm :
      > > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      >
      > > and my DICOMTEST3.dcm :
      > > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      >
      > > How can my file was interpreted as (0002,4955) when the 39da47b7.dcm
      > > was interpreted as (0002,0002) and "UI" ???
      > > All the meta values are the same. The different things are : the value
      > > of date, patient name, and waveform signal.

      >
      > > Is the DICOM waveform file should has the same end of bits of the
      > > file ?

      >
      > > Please advise,
      > > Budhi

      >
      > Hi, I already upload the 4th result file inwww.materikuliah.com/DICOMTEST4.zip
      > (it was zipped).
      > The good thing is the dcm file can be opened by using Dicom
      > Inspector .Net , and I can see the wrong length of VR and the wrong
      > value also. I already fix them based on David's ecg_ow.dcm file and
      > 39da47b7.dcm file.
      >
      > But, again, it can be initialized. I suspect the next wrong thing is
      > in the waveform data syntax.
      >
      > Here is my (5400,1010) "OW" with 0x0000 of reserved and 0xffffffff of
      > length (in C code) :
      >
      > ------- headers -----
      >
      > typedef struct specstruct {
      > unsigned int ge;
      > unsigned char vr[2];
      > unsigned short reserved;
      > unsigned long vl;
      > };
      > struct specstruct metaspec;
      >
      > float date,id,I,II,III,aVR,aVL,aVF,V1,V2,V3,V4,V5,V6;
      >
      > ------- main (bla... bla...) ----------
      >
      > fpecg = fopen("12ecgsample2.txt", "r");
      > fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
      > &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);
      >
      > metaspec.ge = 0x10105400;
      > strcpy(metaspec.vr,"OW");
      > metaspec.reserved = 0x0000;
      > metaspec.vl = 0xffffffff;
      > fwrite(&metaspec,sizeof(metaspec),1,fp);
      >
      > while (!feof(fpecg))
      > {
      > fwrite(&I,sizeof(I),1,fp);
      > fwrite(&II,sizeof(II),1,fp);
      > fwrite(&III,sizeof(III),1,fp);
      > fwrite(&aVR,sizeof(aVR),1,fp);
      > fwrite(&aVL,sizeof(aVL),1,fp);
      > fwrite(&aVF,sizeof(aVF),1,fp);
      > fwrite(&V1,sizeof(V1),1,fp);
      > fwrite(&V2,sizeof(V2),1,fp);
      > fwrite(&V3,sizeof(V3),1,fp);
      > fwrite(&V4,sizeof(V4),1,fp);
      > fwrite(&V5,sizeof(V5),1,fp);
      > fwrite(&V6,sizeof(V6),1,fp);
      >
      > fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
      > &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);
      >
      > }
      >
      > ---------- continue to the end --------------
      >
      > The signal value of each ecg channels are in float numbers %2.6f.
      >
      > Question :
      >
      > 1. Is it correct to write the stream of signals without delimiter
      > between signals ?
      > 2. Should it (the signals) converted first into hex before written
      > into the file ?
      > 3. How to count the group length (0002,0000) value ?
      > 4. Is the (003a,0010) number of waveform samples should be exactly the
      > same with the number of lines of signals in the text file ?
      >
      > Please advise,
      > Budhi


      This is the newest result : www.materikuliah.com/DICOMTEST6.zip

      Based on Dicom Inspector .Net, all tags are OK. The last problem is in
      waveform padding data and waveform data encoding itself.

      1. Should I use delimiter code like "," or " " between successive
      signals ? I know the queue should be C1S1, C2S1,C3S1, ... CnS1,C1S2,
      C2S2, C3S2, ... CnSm. In the supplement 30, it was "with no padding or
      explicit delimitation between successive samples" written. But I have
      tried both, with and without delimitation code, but still not works.

      2. Since my raw data is in float format (i.e. 122.001245 or %3.6f
      format), is it possible to load them directly ? Or should I use only
      decimal number (i.e. 20 or %d format) ?

      Please advise,
      Budhi

  12. Re: HELP : DICOM initialization failure (version 2)

    On Jul 21, 3:29 pm, budhik wrote:
    > On Jul 20, 1:14 am, budhik wrote:
    >
    >
    >
    > > On Jul 19, 8:08 pm, budhik wrote:

    >
    > > > On Jul 19, 4:31 am, David Clunie wrote:

    >
    > > > > budhik wrote:
    > > > > > Based on that, I revised the code and generate new dcm file. This file
    > > > > > shall contains ecg signals from the text file.
    > > > > > You can see the dcm and txt file here (zipped) :www.materikuliah.com/DICOMTEST3.zip
    > > > > > andwww.materikuliah.com/12ecgsample2.zip

    >
    > > > > ...
    > > > > > But still, the dcm file can't be initialized. I don't know why. I
    > > > > > already compared my dcm file with 39da47b7.dcm (downloaded from
    > > > > >http://www.excel-medical.com/waveforms/Waveform.htm), and I found the
    > > > > > meta section seems the same. The only differences are waveform data
    > > > > > and several info like patient name etc.

    >
    > > > > % andump /Users/dclunie/Downloads/DICOMTEST3.dcm
    > > > > (0x0002,0x0000) UL File Meta Information Group Length VR=
      VL=<0x0004> [0x0000008c]
      > > > > (0x0002,0x0001) OB File Meta Information Version VR= VL=<0x0004> [0x01,0x00,0x02,0x00]
      > > > > (0x0002,0x4955) ? VR= VL=<0x2e31001e> [] # skipping ...

      >
      > > > > % dciodvfy /Users/dclunie/Downloads/DICOMTEST3.dcm
      > > > > Warning - Bad File Meta Information Version Attribute - wrong VR, length or not 0x00,0x01
      > > > > (0x0002,0x4955) ? Error - Tag read failed - Implicit VR encoding even though supposed to be explicit
      > > > > Error - Dicom dataset read failed

      >
      > > > Hi,

      >
      > > > I don't understand how the mis-interpretation happened. As we can see,
      > > > the bit values are the same.

      >
      > > > The 39da47b7.dcm :
      > > > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      >
      > > > and my DICOMTEST3.dcm :
      > > > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      >
      > > > How can my file was interpreted as (0002,4955) when the 39da47b7.dcm
      > > > was interpreted as (0002,0002) and "UI" ???
      > > > All the meta values are the same. The different things are : the value
      > > > of date, patient name, and waveform signal.

      >
      > > > Is the DICOM waveform file should has the same end of bits of the
      > > > file ?

      >
      > > > Please advise,
      > > > Budhi

      >
      > > Hi, I already upload the 4th result file inwww.materikuliah.com/DICOMTEST4.zip
      > > (it was zipped).
      > > The good thing is the dcm file can be opened by using Dicom
      > > Inspector .Net , and I can see the wrong length of VR and the wrong
      > > value also. I already fix them based on David's ecg_ow.dcm file and
      > > 39da47b7.dcm file.

      >
      > > But, again, it can be initialized. I suspect the next wrong thing is
      > > in the waveform data syntax.

      >
      > > Here is my (5400,1010) "OW" with 0x0000 of reserved and 0xffffffff of
      > > length (in C code) :

      >
      > > ------- headers -----

      >
      > > typedef struct specstruct {
      > > unsigned int ge;
      > > unsigned char vr[2];
      > > unsigned short reserved;
      > > unsigned long vl;
      > > };
      > > struct specstruct metaspec;

      >
      > > float date,id,I,II,III,aVR,aVL,aVF,V1,V2,V3,V4,V5,V6;

      >
      > > ------- main (bla... bla...) ----------

      >
      > > fpecg = fopen("12ecgsample2.txt", "r");
      > > fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
      > > &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);

      >
      > > metaspec.ge = 0x10105400;
      > > strcpy(metaspec.vr,"OW");
      > > metaspec.reserved = 0x0000;
      > > metaspec.vl = 0xffffffff;
      > > fwrite(&metaspec,sizeof(metaspec),1,fp);

      >
      > > while (!feof(fpecg))
      > > {
      > > fwrite(&I,sizeof(I),1,fp);
      > > fwrite(&II,sizeof(II),1,fp);
      > > fwrite(&III,sizeof(III),1,fp);
      > > fwrite(&aVR,sizeof(aVR),1,fp);
      > > fwrite(&aVL,sizeof(aVL),1,fp);
      > > fwrite(&aVF,sizeof(aVF),1,fp);
      > > fwrite(&V1,sizeof(V1),1,fp);
      > > fwrite(&V2,sizeof(V2),1,fp);
      > > fwrite(&V3,sizeof(V3),1,fp);
      > > fwrite(&V4,sizeof(V4),1,fp);
      > > fwrite(&V5,sizeof(V5),1,fp);
      > > fwrite(&V6,sizeof(V6),1,fp);

      >
      > > fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
      > > &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);

      >
      > > }

      >
      > > ---------- continue to the end --------------

      >
      > > The signal value of each ecg channels are in float numbers %2.6f.

      >
      > > Question :

      >
      > > 1. Is it correct to write the stream of signals without delimiter
      > > between signals ?
      > > 2. Should it (the signals) converted first into hex before written
      > > into the file ?
      > > 3. How to count the group length (0002,0000) value ?
      > > 4. Is the (003a,0010) number of waveform samples should be exactly the
      > > same with the number of lines of signals in the text file ?

      >
      > > Please advise,
      > > Budhi

      >
      > This is the newest result :www.materikuliah.com/DICOMTEST6.zip
      >
      > Based on Dicom Inspector .Net, all tags are OK. The last problem is in
      > waveform padding data and waveform data encoding itself.
      >
      > 1. Should I use delimiter code like "," or " " between successive
      > signals ? I know the queue should be C1S1, C2S1,C3S1, ... CnS1,C1S2,
      > C2S2, C3S2, ... CnSm. In the supplement 30, it was "with no padding or
      > explicit delimitation between successive samples" written. But I have
      > tried both, with and without delimitation code, but still not works.
      >
      > 2. Since my raw data is in float format (i.e. 122.001245 or %3.6f
      > format), is it possible to load them directly ? Or should I use only
      > decimal number (i.e. 20 or %d format) ?
      >
      > Please advise,
      > Budhi


      Just asking, does this forum related with this one :
      HELP : DICOM initialization failure (version 2)
      ??

      Because all records are shown in that URL.

      Budhi

  13. Re: HELP : DICOM initialization failure (version 2)

    On Jul 19, 7:14 pm, budhik wrote:
    > On Jul 19, 8:08 pm, budhik wrote:
    >
    >
    >
    > > On Jul 19, 4:31 am, David Clunie wrote:

    >
    > > > budhik wrote:
    > > > > Based on that, I revised the code and generate new dcm file. This file
    > > > > shall contains ecg signals from the text file.
    > > > > You can see the dcm and txt file here (zipped) :www.materikuliah.com/DICOMTEST3.zip
    > > > > andwww.materikuliah.com/12ecgsample2.zip

    >
    > > > ...
    > > > > But still, the dcm file can't be initialized. I don't know why. I
    > > > > already compared my dcm file with 39da47b7.dcm (downloaded from
    > > > >http://www.excel-medical.com/waveforms/Waveform.htm), and I found the
    > > > > meta section seems the same. The only differences are waveform data
    > > > > and several info like patient name etc.

    >
    > > > % andump /Users/dclunie/Downloads/DICOMTEST3.dcm
    > > > (0x0002,0x0000) UL File Meta Information Group Length VR=
      VL=<0x0004> [0x0000008c]
      > > > (0x0002,0x0001) OB File Meta Information Version VR= VL=<0x0004> [0x01,0x00,0x02,0x00]
      > > > (0x0002,0x4955) ? VR= VL=<0x2e31001e> [] # skipping ...

      >
      > > > % dciodvfy /Users/dclunie/Downloads/DICOMTEST3.dcm
      > > > Warning - Bad File Meta Information Version Attribute - wrong VR, length or not 0x00,0x01
      > > > (0x0002,0x4955) ? Error - Tag read failed - Implicit VR encoding even though supposed to be explicit
      > > > Error - Dicom dataset read failed

      >
      > > Hi,

      >
      > > I don't understand how the mis-interpretation happened. As we can see,
      > > the bit values are the same.

      >
      > > The 39da47b7.dcm :
      > > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      >
      > > and my DICOMTEST3.dcm :
      > > 02 00 01 00 4f 42 00 00 02 00 00 00 00 01 02 00 02 00 55 49

      >
      > > How can my file was interpreted as (0002,4955) when the 39da47b7.dcm
      > > was interpreted as (0002,0002) and "UI" ???
      > > All the meta values are the same. The different things are : the value
      > > of date, patient name, and waveform signal.

      >
      > > Is the DICOM waveform file should has the same end of bits of the
      > > file ?

      >
      > > Please advise,
      > > Budhi

      >
      > Hi, I already upload the 4th result file inwww.materikuliah.com/DICOMTEST4.zip
      > (it was zipped).
      > The good thing is the dcm file can be opened by using Dicom
      > Inspector .Net , and I can see the wrong length of VR and the wrong
      > value also. I already fix them based on David's ecg_ow.dcm file and
      > 39da47b7.dcm file.
      >
      > But, again, it can be initialized. I suspect the next wrong thing is
      > in the waveform data syntax.
      >
      > Here is my (5400,1010) "OW" with 0x0000 of reserved and 0xffffffff of
      > length (in C code) :
      >
      > ------- headers -----
      >
      > typedef struct specstruct {
      > unsigned int ge;
      > unsigned char vr[2];
      > unsigned short reserved;
      > unsigned long vl;
      > };
      > struct specstruct metaspec;


      You cannot do that, you need to differentiate based on VR value:

      7.1.2 DATA ELEMENT STRUCTURE WITH EXPLICIT VR

      You need a special handling for the following VRs:

      ....
      VL32 = OB | OW | OF | SQ | UN | UT, // if( VR & VR_VL32 ) => VR
      has its VL coded over 32bits
      ....

      Do not reuse meta data info from other people
      (1.3.6.1.4.1.6018.2.11111), you need to set it one yourself.

      -Mathieu

  14. Re: HELP : DICOM initialization failure (version 2)

    Each waveform sample is written as a 16 bit binary value.

    The values are concatenated together with no delimiters
    (neither ascii, not the Item delimiters that are used with
    undefined length Pixel Data ... all waveform data requires
    defined length).

    If the values in your text file are floating point values
    that are going to be truncated during conversion to 16 bit
    integers, scale them using Channel Sensitivity (003A,0210).
    E.g., if your floating point data is in mV, multiply the
    values by 1000 before encoding them as 16 bit words, and
    send a value of 0.001 for Channel Sensitivity. E.g., given
    float aVR:

    long aVRvalue = (long)(aVR*1000);

    I mentioned before not depending on the native C data types.

    A call like fwrite(&something,sizeof(something),...) is unsafe
    for two reasons:

    - the number of bytes written is dependent on the compiler and the
    host and target

    - the byte order written is dependent on the target machine

    For each value written you must extract the bytes that you want
    into an array of char in the correct order and write it by hand.

    E.g. given a 16 bit value in long value to be written as a US or
    OW little endian:

    char buffer[2];
    buffer[0]=(char)(value&0xff);
    buffer[1]=(char)((value>>8)&0xff);
    fwrite(buffer,2,...);

    This does depend on char being one byte (always true) and a byte
    always being 8 bits (normally true).

    The value length (in bytes) of the Waveform Data attribute is the
    number of samples * number of leads * the number of bytes per sample
    (which is 2 for OW).

    The group length of the meta-information header is required, so you
    need to add up the number of bytes that you are going to write,
    before writing them.

    David

    budhik wrote:

    >> But, again, it can be initialized. I suspect the next wrong thing is
    >> in the waveform data syntax.
    >>
    >> Here is my (5400,1010) "OW" with 0x0000 of reserved and 0xffffffff of
    >> length (in C code) :
    >>
    >> ------- headers -----
    >>
    >> typedef struct specstruct {
    >> unsigned int ge;
    >> unsigned char vr[2];
    >> unsigned short reserved;
    >> unsigned long vl;
    >> };
    >> struct specstruct metaspec;
    >>
    >> float date,id,I,II,III,aVR,aVL,aVF,V1,V2,V3,V4,V5,V6;
    >>
    >> ------- main (bla... bla...) ----------
    >>
    >> fpecg = fopen("12ecgsample2.txt", "r");
    >> fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
    >> &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);
    >>
    >> metaspec.ge = 0x10105400;
    >> strcpy(metaspec.vr,"OW");
    >> metaspec.reserved = 0x0000;
    >> metaspec.vl = 0xffffffff;
    >> fwrite(&metaspec,sizeof(metaspec),1,fp);
    >>
    >> while (!feof(fpecg))
    >> {
    >> fwrite(&I,sizeof(I),1,fp);
    >> fwrite(&II,sizeof(II),1,fp);
    >> fwrite(&III,sizeof(III),1,fp);
    >> fwrite(&aVR,sizeof(aVR),1,fp);
    >> fwrite(&aVL,sizeof(aVL),1,fp);
    >> fwrite(&aVF,sizeof(aVF),1,fp);
    >> fwrite(&V1,sizeof(V1),1,fp);
    >> fwrite(&V2,sizeof(V2),1,fp);
    >> fwrite(&V3,sizeof(V3),1,fp);
    >> fwrite(&V4,sizeof(V4),1,fp);
    >> fwrite(&V5,sizeof(V5),1,fp);
    >> fwrite(&V6,sizeof(V6),1,fp);
    >>
    >> fscanf(fpecg,"%f %f %f %f %f %f %f %f %f %f %f %f %f %f", &date, &id,
    >> &I, &II, &III, &aVR, &aVL, &aVF, &V1, &V2, &V3, &V4, &V5, &V6);
    >>
    >> }
    >>
    >> ---------- continue to the end --------------
    >>
    >> The signal value of each ecg channels are in float numbers %2.6f.
    >>
    >> Question :
    >>
    >> 1. Is it correct to write the stream of signals without delimiter
    >> between signals ?
    >> 2. Should it (the signals) converted first into hex before written
    >> into the file ?
    >> 3. How to count the group length (0002,0000) value ?
    >> 4. Is the (003a,0010) number of waveform samples should be exactly the
    >> same with the number of lines of signals in the text file ?


  15. Re: HELP : DICOM initialization failure (version 2)

    By the way, how are you going to generate SOP, Series, Study
    and Synchronization Frame of Reference UIDs, if you are doing
    all this by hand without a toolkit ?

  16. Re: HELP : DICOM initialization failure (version 2)

    On Jul 21, 7:09 pm, David Clunie wrote:
    > By the way, how are you going to generate SOP, Series, Study
    > and Synchronization Frame of Reference UIDs, if you are doing
    > all this by hand without a toolkit ?


    @Mathew

    In DICOMTEST6.dcm I already followed the VL32 rules. For
    implementation class uid, I will set myown after the result file
    correctly viewed.

    @David

    I just identified which are the mandatory tags for the 12 lead ECG
    only, and write them one by one in C code.

    By the way, I don't clearly understand with your explanations :

    1. ------ all waveform data requires defined length----- how to do
    that ?

    2. ------- if your floating point data is in mV, multiply the values
    by 1000 before encoding them as 16 bit words, and send a value of
    0.001 for Channel Sensitivity-----
    Yes, my raw data are in millivolt. For example, my data is 2.596498.
    After divided (it's divided, not multiplied right ?) by 1000, it is
    0.002596498 volt. How to convert 0.002596498 into 16 bit words in C ?
    And how to get the channel sensitivity's value by using this result ?

    I'm so sorry for my delayed understanding.

    Budhi

  17. Re: HELP : DICOM initialization failure (version 2)

    On Jul 21, 2:34 pm, budhik wrote:
    > On Jul 21, 7:09 pm, David Clunie wrote:
    >
    > > By the way, how are you going to generate SOP, Series, Study
    > > and Synchronization Frame of Reference UIDs, if you are doing
    > > all this by hand without a toolkit ?

    >
    > @Mathew
    >
    > In DICOMTEST6.dcm I already followed the VL32 rules. For
    > implementation class uid, I will set myown after the result file
    > correctly viewed.


    well then you cannot set an undefined length for VR=OW

    -Mathieu

  18. Re: HELP : DICOM initialization failure (version 2)

    On Jul 21, 2:34 pm, budhik wrote:
    > On Jul 21, 7:09 pm, David Clunie wrote:
    >
    > > By the way, how are you going to generate SOP, Series, Study
    > > and Synchronization Frame of Reference UIDs, if you are doing
    > > all this by hand without a toolkit ?

    >
    > @Mathew
    >
    > In DICOMTEST6.dcm I already followed the VL32 rules. For
    > implementation class uid, I will set myown after the result file
    > correctly viewed.
    >
    > @David
    >
    > I just identified which are the mandatory tags for the 12 lead ECG
    > only, and write them one by one in C code.
    >
    > By the way, I don't clearly understand with your explanations :


    IMHO David is referring -among other- to the UID generation. See PS
    3.5 - 2008

    Section 9 Unique Identifiers (UIDs)

    This is not easy to implement in an open source toolkit (that everyone
    can use), it may be easier for you if you control which particular
    hardware/machine will be running your code.

    -Mathieu

  19. Re: HELP : DICOM initialization failure (version 2)

    On Jul 21, 8:40 pm, Mathieu Malaterre
    wrote:
    > On Jul 21, 2:34 pm, budhik wrote:
    >
    > > On Jul 21, 7:09 pm, David Clunie wrote:

    >
    > > > By the way, how are you going to generate SOP, Series, Study
    > > > and Synchronization Frame of Reference UIDs, if you are doing
    > > > all this by hand without a toolkit ?

    >
    > > @Mathew

    >
    > > In DICOMTEST6.dcm I already followed the VL32 rules. For
    > > implementation class uid, I will set myown after the result file
    > > correctly viewed.

    >
    > well then you cannot set an undefined length for VR=OW
    >
    > -Mathieu


    What do you mean ? Please explain me how to fix it.

    Best Regards,
    Budhi

  20. Re: HELP : DICOM initialization failure (version 2)

    On Jul 21, 2:45 pm, budhik wrote:
    > On Jul 21, 8:40 pm, Mathieu Malaterre
    > wrote:
    >
    >
    >
    > > On Jul 21, 2:34 pm, budhik wrote:

    >
    > > > On Jul 21, 7:09 pm, David Clunie wrote:

    >
    > > > > By the way, how are you going to generate SOP, Series, Study
    > > > > and Synchronization Frame of Reference UIDs, if you are doing
    > > > > all this by hand without a toolkit ?

    >
    > > > @Mathew

    >
    > > > In DICOMTEST6.dcm I already followed the VL32 rules. For
    > > > implementation class uid, I will set myown after the result file
    > > > correctly viewed.

    >
    > > well then you cannot set an undefined length for VR=OW

    >
    > > -Mathieu

    >
    > What do you mean ? Please explain me how to fix it.


    You are inserting a data element: (5400,100a) with VR=OW but at the
    same time VL=u/l (== -1). You cannot do that u/l VL is reserved for
    SQ and UN vr only. Compute the actual length of your binary blob and
    set that length.

    HTH