Series with different SOP classes - Allowed? - DICOM

This is a discussion on Series with different SOP classes - Allowed? - DICOM ; Does DICOM permit SOP Instances with different SOP Classes to be part of the same Series UID? or Do all SOP Instances within a single Series UID have to have the same SOP Class? Also, what about multiple modality codes ...

+ Reply to Thread
Results 1 to 2 of 2

Thread: Series with different SOP classes - Allowed?

  1. Series with different SOP classes - Allowed?

    Does DICOM permit SOP Instances with different SOP Classes to be part
    of the same Series UID?

    or

    Do all SOP Instances within a single Series UID have to have the same
    SOP Class?

    Also, what about multiple modality codes within a Series?

    Thanks!


  2. Re: Series with different SOP classes - Allowed?

    SomeHuman wrote:
    > Does DICOM permit SOP Instances with different SOP Classes to be part
    > of the same Series UID?


    Yes, absolutely.

    > Also, what about multiple modality codes within a Series?


    Not allowed, because Modality (0008,0060) is defined on Series level,
    i.e. needs to be kept constant for all instances of a series.

    Regards,
    Marco Eichelberg
    OFFIS